Refine
H-BRS Bibliography
- yes (66) (remove)
Departments, institutes and facilities
- Institut für Sicherheitsforschung (ISF) (66) (remove)
Document Type
- Article (39)
- Conference Object (22)
- Doctoral Thesis (3)
- Part of a Book (1)
- Patent (1)
Year of publication
Language
- English (66) (remove)
Keywords
- Chemometrics (4)
- DNA typing (3)
- Raman spectroscopy (3)
- Classification (2)
- Cooperative Awareness Message (2)
- Discriminant analysis (2)
- Hyperspectral image (2)
- Intelligent Transport System (2)
- Principal Components Analysis (2)
- Privacy (2)
With the increasing demand for ultrapure water in the pharmaceutical and semiconductor industry, the need for precise measuring instruments for those applications is also growing. One critical parameter of water quality is the amount of total organic carbon (TOC). This work presents a system that uses the advantage of the increased oxidation power achieved with UV/O3 advanced oxidation process (AOP) for TOC measurement in combination with a significant miniaturization compared to the state of the art. The miniaturization is achieved by using polymer-electrolyte membrane (PEM) electrolysis cells for ozone generation in combination with UV-LEDs for irradiation of the measuring solution, as both components are significantly smaller than standard equipment. Conductivity measurement after oxidation is the measuring principle and measurements were carried out in the TOC range between 10 and 1000 ppb TOC. The suitability of the system for TOC measurement is demonstrated using the oxidation by ozonation combined with UV irradiation of defined concentrations of isopropyl alcohol (IPA).
Hydrophilic surface-enhanced Raman spectroscopy (SERS) substrates were prepared by a combination of TiO2-coatings of aluminium plates through a direct titanium tetraisopropoxide (TTIP) coating and drop coated by synthesised gold nanoparticles (AuNPs). Differences between the wettability of the untreated substrates, the slowly dried Ti(OH)4 substrates and calcinated as well as plasma treated TiO2 substrates were analysed by water contact angle (WCA) measurements. The hydrophilic behaviour of the developed substrates helped to improve the distribution of the AuNPs, which reflects in overall higher lateral SERS enhancement. Surface enhancement of the substrates was tested with target molecule rhodamine 6G (R6G) and a fibre-coupled 638 nm Raman spectrometer. Additionally, the morphology of the substrates was characterised using scanning electron microscopy (SEM) and Raman microscopy. The studies showed a reduced influence of the coffee ring effect on the particle distribution, resulting in a more broadly distributed edge region, which increased the spatial reproducibility of the measured SERS signal in the surface-enhanced Raman mapping measurements on mm scale.
Hydrogen‐Bonded Cholesteric Liquid Crystals—A Modular Approach Toward Responsive Photonic Materials
(2022)
A supramolecular approach for photonic materials based on hydrogen-bonded cholesteric liquid crystals is presented. The modular toolbox of low-molecular-weight hydrogen-bond donors and acceptors provides a simple route toward liquid crystalline materials with tailor-made thermal and photonic properties. Initial studies reveal broad application potential of the liquid crystalline thin films for chemo- and thermosensing. The chemosensing performance is based on the interruption of the intermolecular forces between the donor and acceptor moieties by interference with halogen-bond donors. Future studies will expand the scope of analytes and sensing in aqueous media. In addition, the implementation of the reported materials in additive manufacturing and printed photonic devices is planned.
Polymer fibers with liquid crystals (LCs) in the core have potential as autonomous sensors of airborne volatile organic compounds (VOCs), with a high surface-to-volume ratio enabling fast and sensitive response and an attractive non-woven textile form factor. We demonstrate their ability to continuously and quantitatively measure the concentration of toluene, cyclohexane, and isopropanol as representative VOCs, via the impact of each VOC on the LC birefringence. The response is fully reversible and repeatable over several cycles, the response time can be as low as seconds, and high sensitivity is achieved when the operating temperature is near the LC-isotropic transition temperature. We propose that a broad operating temperature range can be realized by combining fibers with different LC mixtures, yielding autonomous VOC sensors suitable for integration in apparel or in furniture that can compete with existing consumer-grade electronic VOC sensors in terms of sensitivity and response speed.
Intimate swabs taken for examination in sexual assault cases typically yield mixtures of sperm and epithelial cell types. While powerful, differential extraction protocols to overcome such cell type mixtures by separate lysis of epithelial cells and spermatozoa can still prove ineffective, in particular if only few sperm cells are present or if swabs contain sperm from more than one individual leading to complex low level DNA mixtures. A means to avoid such mixtures consists in the analysis of single micromanipulated sperm cells. However, the quantity of DNA from single sperm cells is not sufficient for conventional STR analysis. Here, we describe a simple method for micromanipulating individual sperm cells from intimate swabs and show that whole genome amplification can generate sufficient amounts of DNA from single cells for subsequent DNA profiling. We recovered over 80% of alleles of haploid autosomal STR profiles from the majority of individual sperm cells. Furthermore, we demonstrate that in mixtures of sperm from two contributors, Y-STR and X-STR profiles of individual sperm cells can be used to sort the haploid autosomal profiles to develop the diploid consensus STR profiles of the individual donors. Finally, by analysing single sperm cells from mock sexual assault swabs with one or two sperm donors, we showed that our protocols enabled the identification of the unknown male contributors.
Because the robust and rapid determination of spoilage microorganisms is becoming increasingly important in industry, the use of IR microspectroscopy, and the establishment of robust and versatile chemometric models for data processing and classification, is gaining importance. To further improve the chemometric models, bacterial stress responses were induced, to study the effect on the IR spectra and to improve the chemometric model. Thus, in this work, nine important food-relevant microorganisms were subjected to eight stress conditions, besides the regular culturing as a reference. Spectral changes compared to normal growth conditions without stressors were found in the spectral regions of 900–1500 cm−1 and 1500–1700 cm−1. These differences might stem from changes in the protein secondary structure, exopolymer production, and concentration of nucleic acids, lipids, and polysaccharides. As a result, a model for the discrimination of the studied microorganisms at the genus, species and strain level was established, with an accuracy of 96.6%. This was achieved despite the inclusion of various stress conditions and times after incubation of the bacteria. In addition, a model was developed for each individual microorganism, to separate each stress condition or regular treatment with 100% accuracy.
A series of reactive binaphthyl‐diimine‐based dopants is prepared and investigated with respect to their potential for the chiral induction of structural coloration in nematic liquid crystal mixture E7 and the selective photonic sensing of nitrogen dioxide (NO2). Studies of the helical twisting power (HTP) in 4‐cyano‐4′‐pentylbiphenyl (5CB) reveal HTP values as high as 375 µm‐1 and the tremendous impact of structural compatibility and changes of the dihedral binaphthyl angle on the efficiency of the chiral transfer. Detailed investigation of the sensing capabilities of the systems reveals an extraordinarily high selectivity for NO2 and a response to concentrations as low as 100 ppm. The systems show a direct response to the analyte gas leading to a concentration‐dependent shift of the reflectance wavelength of up to several hundred nanometers. Incorporation of copper ions remarkably improves the sensor's properties in terms of sensitivity and selectivity, enabling the tailored tweaking of the system's properties.
The simultaneous operation of multiple different semiconducting metal oxide (MOX) gas sensors is demanding for the readout circuitry. The challenge results from the strongly varying signal intensities of the various sensor types to the target gas. While some sensors change their resistance only slightly, other types can react with a resistive change over a range of several decades. Therefore, a suitable readout circuit has to be able to capture all these resistive variations, requiring it to have a very large dynamic range. This work presents a compact embedded system that provides a full, high range input interface (readout and heater management) for MOX sensor operation. The system is modular and consists of a central mainboard that holds up to eight sensor-modules, each capable of supporting up to two MOX sensors, therefore supporting a total maximum of 16 different sensors. Its wide input range is archived using the resistance-to-time measurement method. The system is solely built with commercial off-the-shelf components and tested over a range spanning from 100Ω to 5 GΩ (9.7 decades) with an average measurement error of 0.27% and a maximum error of 2.11%. The heater management uses a well-tested power-circuit and supports multiple modes of operation, hence enabling the system to be used in highly automated measurement applications. The experimental part of this work presents the results of an exemplary screening of 16 sensors, which was performed to evaluate the system’s performance.